Mechanical Pipettes (Adjustable and Fixed Volume)
Calibration
Please visit DRAGONLAB website for online calibration at www.dragon-lab.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation. We will support and help you to achieve consistently excellent results. |
MicroPette Single-channel Adjustable Volume Pipettes | |||||||
Cat No. | Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | ||
% | μl | % | μl | ||||
712111010000 | 0.1-2.5μl | 0.05μl | 2.5 | 2.50% | 0.0625 | 2.00% | 0.05 |
1.25 | 3.00% | 0.0375 | 3.00% | 0.0375 | |||
0.25 | 12.00% | 0.03 | 6.00% | 0.015 | |||
712111040000 | 0.5-10μl | 0.1μl | 10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.075 | 1.50% | 0.075 | |||
1 | 2.50% | 0.025 | 1.50% | 0.015< |