MB-530 Full-Automatic Micro-Plate Reader
  It  is controlled by computer to complete Enzyme Linked Immunosorbent Assay (ELISA).  Read 48 well and 96 well microplate, analyse and report, which are widely used  in clinical diagnostic laboratories, centers for disease control and prevention,  animal and plant quarantine, animal husbandry and veterinary epidemic prevention  stations, biotechnological industries, food industries, environmental science,  agricultural scientific research and other academic organizations. 

Features
  It is controlled by external  computer to complete edit, test, save, print efficiently and conveniently.  
  8 channels zero dispersion monochromatic optical fiber testing system,  automatic well center     positioning.
  High-precision rail transmission,  enclosing optical system, test 96 wells in 3 seconds.
  System self-check and  diagnosis function for optical paths and mechanical parts.
  Lamp’s automatic  switch, energy-efficient design extends the life of light source and makes easy  to     change lamp.
  Shaking function, adjustable shaking speed and  time.
  New Windows graphical user interface, operation system in  English.
  Specimen information records: lists and cards, directly input  specimen No., Patient Name, etc.
  Specimen No., OD value, result, S/CO value  and term parameters can be shown in the same     screen.
  Quick tagging  blank, specimen, positive control, quality control, multi-valued  comparison.
  Arrange 96 well microplate horizontally or longitudinally, set  up optionally the start bit and stop bit in   testing, autocoding.
  Suitable  for self contrast, rows and columns subtraction, input judgment formula  freely.
  Qualitative and quantitive parameters, the storable standard curve  can be used directly.
  Multiterm can be tested on one microplate up to 12  terms.
  Report according to microplate No. or specimen No., input  experimental terms and results in     batches.
  Inquiry and statistic  analysis can be done according to the specimen No. and name, automatically      track and analyse changing trends in data.
  Test data can be saved in MS  Excel for further analysis.
Main Technical Parameters
  Light  Source: 6V, 10W tungsten halogen lamp, it can work over 5000  hours.
  Wavelength Range: 400~750  nm.
  Optical Filters: 4 standard filters: 405, 450, 492,  630 nm, maximum 8 filters.
  Measurement Methods: single  wavelength, dual wavelength, two point method, dynamic assay,     multiple  wavelengths.
  Blank Ways: single well blank, multiple wells  blank, rows and columns blank.
  Calculation Method:  absorbance, single point calibration, linear regression, quadratic  curve, four     parameters equation.
  Measurement Speed:  single wavelength < 3 seconds/96 wells, dual wavelength < 9  seconds/96     wells.
  Measurement Range:  0.0000-4.5000Abs.
  Resolution:  0.0001A.
  Linearity: ±0.5% or < 0.025  A.
  Repeatability Difference: < 0.1% or  ±0.0025A.
  Rows and Columns Difference: < 0.01  A.
  Display: controlled by external computer, shows whole  microplate results and calibration curve.
  Save Information:  more than 300 test programs, results of 50 pieces of 96 well, 100000  test     results.
  Interface: 4 USB two-way  communication interface, 1 RS232, 1 VGA , 1 Keyboard connector, 1     mouse  connector.
  Environment: power supply 12V DC (100V-240V  wide input voltage range); temperature: 5℃-40℃;   Humidity:  10%-90%.